Reversed-phase Liquid chromatography (RP-LC) is a mode of liquid chromatography in which non-polar stationary phase and polar mobile phases are used for the separation of organic comounds. The vast majority of separations and analyses using High Performance Liquid Chromatography-HPLC in recent years are done using the Reversed Phase mode. In the Reversed Phase mode, the sample components are retained in the system, the more hydrophobic they are.
The factors affecting the retention and separation of solutes in the Reversed Phase chromatographic system are as follows:
a. The chemical nature of the stationary phase, i.e., the ligands bonded on its surface, as well as their bonding density, namely the extent of their coverage.
b. The composition of the mobile phase. Type of the bulk solvents whose mixtures affect the polarity of the mobile phase, hence the name modifier for a solvent added to affect the polarity of the mobile phase.
C. Additives, such as buffers, affect the pH of the mobile phase, which affect the ionization state of the solutes and their polarity.
In order to retain the organic components in mixtures, the stationary phases, packed within columns, consist of a hydrophobic substrates, bonded to the surface of porous silica-gel particles in various geometries (spheric, irregular), at different diameters (sub-2, 3, 5, 7, 10 um), with varying pore diameters (60, 100, 150, 300, A). The particle's surface is covered by chemically bonded hydrocarbons, such as C3, C4, C8, C18 and more. The longer the hydrocarbon associated with the stationary phase, the longer the sample components will be retained. Some stationary phases are also made of hydrophobic polymeric particles, or hybridized silica-organic groups particles, for method in which mobile phases at extreme pH are used. Most current methods of separation of biomedical materials use C-18 columns, sometimes called by trade names, such as ODS (octadecylsilane) or RP-18.
The mobile phases are mixtures of water and polar organic solvents, the vast majority of which are methanol and acetonitril. These mixtures usually contain various additives such as buffers (acetate, phosphate, citrate), and/or surfactants (alkyl amines or alkyl sulfonates) and/or special additives (EDTA). The goal of using supplements of one kind or another is to increase efficiency and/or selectivity, and control solute retention.
Stationary phases
The history and evolution of Reversed Phase stationary phases in described in details in an article by Majors, Dolan, Carr and Snyder
In the 1970s, most liquid chromatography runs were performed using solid particles as the stationary phases, made of unmodified silica-gel or alumina. This type of technique is now referred to as normal-phase chromatography. Since the stationary phase is hydrophilic in this technique, and the mobile phase were a-polar, consisted of non-polar organic solvents such as hexane and heptane, biomolecules with hydrophilic properties in the sample had high affinity to the stationary phase, therefore they adsorb to it strongly. Moreover, they were not dissolved easily in the mobile phase solvents. At the same time hydrophobic molecules experience less affinity to the polar stationary phase, and elute through it early with not enough retention. This was the reasons why during the 1970's the silica based particles were treated with hydrocarbons, immobilized or bonded on their surface, and the mobile phases were switched to aqueous and polar in nature, to accommodate biomedical substances.
The use of a hydrophobic stationary phase and polar mobile phases is essentially the reverse of normal phase chromatography, since the polarity of the mobile and stationary phases have been inverted – hence the term reversed-phase chromatography. As a result, hydrophobic molecules in the polar mobile phase tend to adsorb to the hydrophobic stationary phase, and hydrophilic molecules in the sample pass through the column and are eluted first. Hydrophobic molecules can be eluted from the column by decreasing the polarity of the mobile phase using an organic (non-polar) solvent, which reduces hydrophobic interactions. The more hydrophobic the molecule, the more strongly it will bind to the stationary phase, and the higher the concentration of organic solvent that will be required to elute the molecule.